COMMUNICATION INTERFACE

71

Parameter /
Return Parameter

Sets the voltage range in volts. The unit
will automatically choose the closest
range that is programmed.
DC Range : 0.0000001 ~ 1050
AC Range : 0.0000001 ~ 600

AUTO

Sets the range to AUTO.

Example

VOLT:RANG AUTO
Sets the voltage range to auto.

Note

The range that is returned is the base voltage unit. See
the table below:

Unit

Voltage Range

1000

1000VDC

600

600ACV

100

200V

10

20V

1

2V

0.1

200mV

[SENSe:]VOLTage:DC:AVERage:COUNt

Set

Query

Description

This command will set or return the average count
setting for DC voltage.

Syntax
Query Syntax

[SENSe:]VOLTage:DC:AVERage:COUNt
[SENSe:]VOLTage:DC:AVERage:COUNt?

Parameter /
Return Parameter

The average count setting for DC voltage.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100

Query Example

VOLT:DC:AVER:COUN?
>10
The average count setting for DC voltage is 10.