82

Syntax

*ESE (0~255)

Query Syntax

*ESE?

Parameter /
Return Parameter

0~255: Indicates the bit weight of the
Standard Event Enable Register.

Query Example

*ESE 189
Enables all bits except for bit 1 and 6.

*ESR?

Query

Description

Queries the Standard Event Register.

Bit

Bit weight Description

0

1

Operation Complete

1

2

Not used

2

4

Query Error

3

8

Device Error

4

16

Execution Error

5

32

Command Error

6

64

Not used

7

128

Power On

Query Syntax

*ESR?

Parameter

0~255: Indicates the bit weight of the
Standard Event Register.

Query Example

*ESR?
>32
Indicates a command error was encountered.

*SRE

Set

Query

Description

Returns or sets the contents of the Service Request
Enable Register.

Bit

Bit weight Description

0

1

Not used

1

2

Not used