COMMUNICATION INTERFACE

63

Example

CONF:CURR 20
Sets the current mode to DC and the range to 30A

Example

CONF:CURR
Sets the current mode to DC. The range is not
changed.

CONFigure:CURRent:AC

Set

Description

This command will set current mode to AC and set
the range. If the range is not specified, then it will
not change.

Syntax

CONFigure:CURRent:AC [ | AUTO]

Parameter

Current range : 0.00000001~305
Current range. The unit will
automatically be set to the closest range.

AUTO

Autorange; Only applicable for the ≤3A
ranges.
Autorange is not supported for the 30A
and 300A ranges.

Example

CONF:CURR:AC 100
Sets the current mode to AC and the range to 300A.

Example

CONF:CURR:AC
Sets the current mode to AC. The range is not
changed.

CONFigure:VOLTage

Query

Description

The CONFigure:VOLTage query will return the
voltage mode and the voltage range unit.

Query Syntax

CONFigure:VOLTage?

Return Parameter

Returns the voltage mode and range unit.