COMMUNICATION INTERFACE

69

Sense Commands

[SENSe:]CURRent:RANGe .......................................... 69

[SENSe:]CURRent:DC:AVERage:COUNt .................... 70

[SENSe:]CURRent:AC:AVERage:COUNt .................... 70

[SENSe:]VOLTage:RANGe .......................................... 70

[SENSe:]VOLTage:DC:AVERage:COUNt .................... 71

[SENSe:]VOLTage:AC:AVERage:COUNt .................... 72

[SENSe:]CURRent:RANGe

Set

Query

Description

Sets or queries the current range.

Syntax
Query Syntax

[SENSe:]CURRent:RANGe {|AUTO}
[SENSe:]CURRent:RANGe?

Parameter /
Return Parameter

Current range : 0.00000001~305
Sets the current range in amps. The unit
will automatically choose the closest
range that is programmed.

AUTO

Sets the range to AUTO; Only applicable
for the ≤3A ranges.
Autorange is not supported for the 30A
and 300A ranges.

Example

CURR:RANG AUTO
Sets the current range to AUTO.

Note

The range that is returned is the base unit. See the
table below:

Unit

Current Range

100

300A

10

30A

1

3A

.1

300mA

.01

30mA