beautypg.com

Status:operation:enable, Status:operation[:event – GW Instek PCS-1000 User Manual

Page 78

background image

PCS-1000 User Manual

78

STATus:OPERation:ENABle

Set

Query

Description

Returns or sets the contents of the Standard
Operation Enable Register.

Bit

Bit weight Description

0

1

Calibrating

1~3

~

Not used

4

16

Measuring

5~7

~

Not used

8

256

Config Change

9~15

~

Not used

Syntax

STATus:OPERation:ENABle (0~65535)

Query Syntax

STATus:OPERation:ENABle?

Parameter /
Return Parameter

0~65535: Indicates the bit weight of the
Standard Operation Enable Register.

Query Example

STAT:OPER:ENAB 273
Enables bit 0, 4 and 8 of the Standard Operation
Enable Register.

STATus:OPERation[:EVENt]

Query

Description

Returns the contents of the Standard Operation
Event Register.

Bit

Bit weight Description

0

1

Calibrating

1~3

~

Not used

4

16

Measuring

5~7

~

Not used

8

256

Config Change

9~15

~

Not used

Query Syntax

STATus:OPERation[:EVENt]?

Return Parameter

0~65535: Returns the bit weight of the
Standard Operation Event Register.