beautypg.com

Source[1]:frequency:stop – GW Instek AFG-2000 User Manual User Manual

Page 127

background image

REMOTE INTERFACE

125

Example

SOUR1:FREQ:STAR +2.0000E+03

Sets the start frequency to 2kHz.

Query Syntax

SOURce[1]:FREQuency:STARt? [MINimum|
MAXimum]

Return Parameter

Returns the start frequency in Hz.

Query Example

SOUR1:FREQ:STAR? MAX
>+2.0000E+07

Returns the maximum start frequency allowed.

SOURce[1]:FREQuency:STOP

Set

Query

Description

Sets the stop frequency of the sweep. 1 kHz is the
default start frequency.

Note

To sweep up set the stop frequency higher than the
start frequency. Set the stop frequency lower than
the start frequency to sweep down.

Syntax

SOURce[1]:FREQuency:STOP
{|MINimum|MAXimum}

Parameter

0.1Hz ~ 25MHz*

0.1Hz ~ 1MHz (Ramp)

*AFG-2112 limited to 12MHz, AFG-2105 limited to
5MHz.

Query Example

SOUR1:FREQ:STOP +2.0000E+03

Sets the stop frequency to 2kHz.

Query Syntax

SOURce[1]:FREQuency:STOP? [MINimum|
MAXimum]

Return Parameter

Returns the stop frequency in Hz.

Example

SOUR1:FREQ:STOP? MAX
>+2.0000E+07

Returns the maximum stop frequency allowed.