Source[1]:fm:deviation – GW Instek AFG-2000 User Manual User Manual
Page 119

REMOTE INTERFACE
117
Return Parameter 
Returns the frequency in Hz.
Query Example
SOUR1:FM:INT:FREQ? MAX 
>+2.0000E+04
Returns the maximum frequency allowed.
SOURce[1]:FM:DEViation
Set
Query
Description
Sets or queries the peak frequency deviation of the 
modulating waveform from the carrier waveform. 
The default peak deviation is 100Hz. 
The frequency deviation of external sources is 
controlled using the ±5V MOD INPUT terminal on 
the rear panel. A positive signal (>0~+5V) will 
increase the deviation (up to the set frequency 
deviation), whilst a negative voltage will reduce 
the deviation. 
Note
The relationship of peak deviation to modulating 
frequency and carrier frequency is shown below. 
Peak deviation = modulating frequency – carrier 
frequency. 
The carrier frequency must be greater than or 
equal to the peak deviation frequency. The sum of 
the deviation and carrier frequency must not 
exceed the maximum frequency for a specific 
carrier shape + 1kHz. If an out of range deviation 
is set for any of the above conditions, the deviation 
will be automatically adjusted to the maximum 
value allowed and an ―out of range‖ error will be 
generated. 
For square wave carrier waveforms, the deviation 
may cause the duty cycle frequency boundary to 
be exceeded. In these conditions the duty cycle will 
be adjusted to the maximum allowed and a ―-221‖ 
error will be generated. 