AFG-2000 Series User Manual

98


This means that the magnitude of the DC offset is
determined by the output amplitude.

If the specified DC offset is out of range, the
maximum/minimum offset will be set instead. A
―-222‖ error will be generated from the remote
terminal.

SOURce[1]:APPLy:SINusoid

Set

Description

Outputs a sine wave when the command has
executed. Frequency, amplitude and offset can also
be set.

Syntax

SOURce[1]:APPLy:SINusoid [
[, [,] ]]

Parameter

0.1Hz~25MHz*

1mV~10Vpp (50Ω)

-5V ~ +5V (50Ω)

*AFG-2005/2105 limited to 5MHz, AFG-2012/2112
limited to 12MHz.

Example

SOURce1:APPL:SIN MAX, 3.0, -2.5

Outputs a 3Vpp sine wave at 25MHz (max
frequency) with a -2.5V offset.

SOURce[1]:APPLy:SQUare

Set

Description

Outputs a square wave when the command has
executed. Frequency, amplitude and offset can also
be set. The duty cycle is fixed to 50%.

Syntax

SOURce[1]:APPLy:SQUare [ [,
[,] ]]

Parameter

0.1Hz ~ 25MHz*

<

amplitude

>

1mV~10V (50Ω)