beautypg.com

Ref:locate, Ref:save – GW Instek GDS-1000A-U Series Programming Manual User Manual

Page 73

background image

COMMAND DETAILS

73

:REF:LOCate

Set

Query

Description

Moves or returns the position of a reference
waveform.
Same as: Save/Recall key → F5 → Variable knob

Syntax

< Long >

< Short >

:ref:locate
:ref:locate?

:ref:loc
:ref:loc?

Parameter

Reference

Position

1

A

–100 to +100

2

B

Note

Before using this command, turn on a reference
waveform. See the example below.

Example

:ref1:display 1
:ref1:locate 0

Turns on the reference
waveform A and move
it to ±0 position

:REF:SAVe

Set

Description

Saves an input signal as a reference waveform.
Same as: Save/Recall key (save) → F2 → F2 → F3

Syntax

< Long >

< Short >

:ref:save

:refsav

Parameter

Reference

Source

1

A

1

Channel 1

2

B

2

Channel 2

3

Math

Example

:ref1:save 1

Saves the Channel 1
signal as the reference
waveform A