beautypg.com

Seed Hawk AIRCART 2013 User Manual

Page 57

background image

SEED HAWK

®

 AIRCART OPERATOR’S MANUAL                                                                              52 

7.3.4

 GROUND SPEED CALIBRATION 


All new systems require a ground speed calibration to ensure accurate area totals and accurate 
ground speed readings. 
 
NOTE: During the calibration the monitor is counting 
the  number  of  pulses  produced  from  the  drill 
mounted sensor.  The monitor will display the pulses 
counted  in  400  feet.  Choosing  this  item  will  allow 
you to calibrate the ground speed sensor to a known 
distance.  The  default  distance  is  400  feet  (122 
meters) but this can be set by the user. To perform a 
ground speed calibration, determine the distance to 
be used and measure out a course.  

 

Procedure: 
 
1.  To  complete  the  calibration,  measure  a  course 
400  feet  (122  meters)  long  preferably  on  level 
ground  with  a  start  and  finish  point.    Line  up 
tractor at the Start Point.  The drill must be in the 
down position throughout the procedure. 
 
2.  On  the  monitor,  verify  that  the  calibration 
length  is  400  feet  (122  meters).    If  not,  please 
change so that it reads 400 feet (122 meters). 
 
3. Once 400 feet (122 meters) is selected, highlight and select Continue. 
 
4.  The  monitor  now  displays  “Start  Driving”  and 
shows  a  Wheel  Pulses  Received  window.    You 
should  see  this  number  increase  as  you  drive  the 
400 foot (122 meter) course. 
 
5.  Once  at  the  end  of  the  400  foot  (122  meter) 
course,  stop  the  tractor,  highlight  the  complete 
button and press Enter. 
 
6. Your Ground Speed is now calibrated.