ଙଝ ଝ – Yokogawa IR400 User Manual
Page 90
7 - 9
IM 11G02N01-01E
7.4 Inspection and maintenance of limited service-life components
The analyzer uses limited-life components. The recommended replacement periods are listed in
the below table.
(1) Limited service-life components are those which wear out or for which failure is presumed
within five years under normal operating or storage conditions. Components with more than
five years of service life are the exception.
(2) The following table only involves the recommended periods for conducting preventive main-
tenance for limited service-life components; these periods do not guarantee that accidental
failures will not occur.
(3) The recommended replacement periods are tentative and depend on operating conditions.
(4) The recommended replacement periods may vary depending on the field date.
1. Light source
Recommended replacement period: Every five years (*1)
2. Sector motor
Recommended replacement period: Every two years (*1)
3. Detector without
Recommended replacement period: Every five years (*1)
O
2
sensor
4. O-ring
Recommended replacement period: Every two years,
inspect it when cleaning of measuring cell. See Sec. 7.3
5. LCD panel
Recommended replacement period: Every five years (*1)
6. Power supply
Recommended replacement period: Every five years (*1)
7. Measuring cell
Set up an appropriate maintenance period
( by refering to the check results)
8. Reference cell
Recommended replacement period: Every five years (*1)
Checkpoint
Infrared gas analyzer
Check and
maintenance items
Procedure and criteria
Recommended check and
maintenance periods
Every five years
Every two years
Once a year
Every six months
Every three months
Once a month
Once a week
Routine
(*) In the check and maintenance columns, place a check mark (
᭺
) for check and confirmation work,
a dark star (
ଙ
) for replacement, and a white star (
ଝ
) for parts preparation for preventive maintenance.
(*1) These are seviceperson’s work, contact our sevicepersons.
ଙ
ଙ
ଙ
ଙ
ଝ
ଝ
᭺
᭺
T7.5E.eps
ଝ
᭹
Precautions to be taken while checking
(1) When handling reference gas (during calibration), carefully read the reference-gas instruc-
tion manual to use the gas correcrtly. In particular, special attention must be taken in han-
dling carbon monoxide gases; otherwise, you may suffer from gas poisoning.
(2) During maintenance checks, be sure to keep the near fan on. If any gas leaks, you may suffer
from gas poisoning.
(3) When replacing the analyzer gas filter or conducting maintenance service of the washer,
completely shut the calibration-gas valve. Otherwise, you may suffer from gas poisoning.