(1) Press the
ENT
key in a state where the
cursor is placed next to “Flow Time,” and
the flow time setting screen shown at right
appears.
(2) On the flow time setting screen that
appears, move the cursor to the gas you
want to change the setting by pressing the
or the
key, and then press the
ENT
key.
(3) The highlighted value can be changed.
Change the value by pressing the
or
the
key, and then move the cursor to
the right by pressing the
key.
(4) After changing the value, press the
ENT
key.
(5) Press the
ESC
key to return to the
automatic calibration setting screen.
Note) Only the Chs used are displayed on
this screen. The Ex. time is the
output signal hold extension time
after the completion of calibration.
It is valid only when the hold setting
is set to “ON.” The Ex. time set
here is also the hold extension time
at the time of manual calibration.
MODE
(
)
ENT
ENT
End of Gas Flow Time Setting