6 - 1

IM 11G02N01-01E

MODE

(

)

ENT

Range switch

or previous screen

6.1 Switch of range

6.1.1 Setting of range switch mode

Set the range switch mode as follows.

(1) Press the

MODE

key in measurement mode to

display the User mode screen.

(2) Move the cursor to “Switch Ranges” and

press the

ENT

key.

(3) The “Channel Selection” screen appears.

Move the cursor by pressing the

or

the

key on the channel selection

screen that appears, and select Ch
(component).

(4) Then press the

ENT

key.

(5) Selected range switch mode is highlighted.

Press the

or the

key to select a

desired switch mode.

(6) Then press the

ENT

key to confirm the

selection.
If “MR” is selected, the cursor moves to
“Range Switch.”

MR: Select a desired range on this screen.
RR: Select a desired range according to

the remote range switch contact input.

AR: Automatically switched from Range 1

to Range 2 when the measured concen-
tration exceeds 90% of Range 1.
Automatically switched from Range 2
to Range 1 when the measured concen-
tration becomes smaller than 80% of
Range 1.

* Operation set for each Ch only can be

performed.

Description of setting

ENT

(

)

ENT