2 forced run/stop of calibration, 1 execution of auto calibration*only once – Yokogawa IR400 User Manual
Page 56
6 - 17
IM 11G02N01-01E
(
)
ENT
(
)
ENT
6.4.2 Forced run/stop of auto calibration
Auto calibration can be performed just once or forcibly stopped while the calibration is
performed.
6.4.2.1 Execution of auto calibration (only once)
(1) Display the User mode screen. Move the
cursor to “Setting of Auto Calibration”
by pressing the
or the
key, and
then press the
ENT
key.
(2) In the “Setting of Auto Calibration” item
selection screen that appears, point the
cursor to “Auto Calibration Run” by
pressing the
or
key. Press the
ENT
key.
(3) “Run” is highlighted, displaying a mes-
sage to confirm the execution of auto
calibration. Press the
ENT
key to execute
the auto calibration forcibly, and press
the
ESC
key to cancel.