2 setting of manual zero calibration – Yokogawa IR400 User Manual
Page 44
6 - 5
IM 11G02N01-01E
(
)
ENT
(
)
ENT
MODE
(
)
ENT
(1) During measurement, press the
MODE
key
to display the User mode.
(2) Point the cursor to “Calibration Param-
eters” by pressing the
or
key. Press
the
ENT
key.
(3) In the “Calibration Parameters” screen that
appears, point the cursor to “About ZERO
Calibration” by pressing the
or
key.
Press the
ENT
key.
(4) In the “Manual ZERO Calibration Ch
Selection” screen that appears, point the
cursor to Ch (component) you want to set
by using the
or
key. Press the
ENT
key.
6.2.2 Setting of manual zero calibration
When zero calibration is made manually, set either all measurement components should be
calibrated simultaneously or each component should be calibrated while selecting one by one.