beautypg.com

Yokogawa IR400 User Manual

Page 75

background image

6 - 36

IM 11G02N01-01E

• Moisture interference adjustment screen

For details of this item, refer to “Item 7.3.4
Moisture interference adjustment method.”

Move the cursor to a desired Ch (compo-

nent) by pressing the

or the

key, and

then press the

ENT

key, and the selected value

at right is highlighted.

Check that the gas for moisture interference
compensation is flowing, change the mois-
ture interference compensation coefficient

using the

or the

key, adjust the value

at left so that it becomes near zero, and then

press the

ENT

key to log moisture interfer-

ence compensation value.

In values on the left side of screen, the moisture
interference for each component is already offset.
The figures at right are interference compensation
coefficients.

Description of moisture interference
adjustment screen

Caution

Since an interference compensation detector
is not provided if the 1st range is beyond 0 to
10 vol%, no interference adjustment can be
performed (no need).

ENT