• Moisture interference adjustment screen
For details of this item, refer to “Item 7.3.4
Moisture interference adjustment method.”
Move the cursor to a desired Ch (compo-
nent) by pressing the
or the
key, and
then press the
ENT
key, and the selected value
at right is highlighted.
Check that the gas for moisture interference
compensation is flowing, change the mois-
ture interference compensation coefficient
using the
or the
key, adjust the value
at left so that it becomes near zero, and then
press the
ENT
key to log moisture interfer-
ence compensation value.
In values on the left side of screen, the moisture
interference for each component is already offset.
The figures at right are interference compensation
coefficients.
Description of moisture interference
adjustment screen
Caution
Since an interference compensation detector
is not provided if the 1st range is beyond 0 to
10 vol%, no interference adjustment can be
performed (no need).