Yokogawa IR400 User Manual

Page 74

background image

6 - 35

IM 11G02N01-01E

• Calibration Log screen

Past calibration history.
Sensor input value, concentration value, and
the date when zero/span calibration is performed
are logged. The 10 newest calibration data is
logged by each component.
Move the cursor to Clear Calibration Log and
press the key, and the calibration log is
cleared completely.

Z1 : Zero calibration (Z) of Range 1
S1 : Span calibration (S) of Range 1
M

: Value of measuring detector at the time

of calibration

C

: Value of the interference compensation

detector at the time of calibration

Con : Concentration value displayed before

calibration

Description of Calibration Log screen

ENT

If the following operation is maladjusted, the measurement may be adversely and excessively
affected. Carry out the operation with utmost attention.

Caution

. Optical adjustment screen

For details of this item, refer to “Item 7.3.3
Optical zero adjustment method”.

Press

ENT

key and turn ON the solenoid

valve signal for each calibration gas

by using the

or

key.

ENT