3 alarm setting, 1 setting of alarm values – Yokogawa IR400 User Manual
Page 50
6 - 11
IM 11G02N01-01E
(
)
ENT
MODE
(
)
ENT
Note
Set the values so that H-limit value > L-limit
value and that (H-limit value
-
L-limit value)
> hysteresis.
6.3 Alarm setting
6.3.1 Setting of alarm values
The High/Low limit alarm output setting for the measured concentration and power off alarm
(alarm 6 only) setting can be made during measurement. Arbitrary 6 alarm contact outputs can be
used.
To change alarm setting, set the alarm ON/OFF setting to OFF, and then change the value.
(1) During measurement, press the
MODE
key
to display the User mode.
(2) Point the cursor to “Alarm Setting” by
pressing the
or
key. Press the
ENT
key.
(3) After the alarm No. selection screen has
appeared, point the cursor to the Alarm
No. you want to set by pressing
or
key .
Press the
ENT
key.
(4) After the alarm item selection screen has
appeared, operate the
or
key until
the cursor is aligned with a desired item
and press the
ENT
key.
(
)