7- 8

IM 11G02N01-01E

7.3.5 Replacement of fuse on analyzer unit

Note) Before the following work, be sure to remove the cause of blown fuse (short, etc) and

repair.

(1) Turn OFF the main power supply SW of

the analyzer.

(2) Loosen setscrews (4 pcs.) from the top of

the analyzer and remove the top cover.

(3) Turn the fuse holder (shown in Fig. 7-3)

counterclockwise and put it out, and the cap

will be removed. Remove the fuse

(250 V AC/3 A) from the holder. Replace it

with a new one .

(4) Mount the fuse holder cap and cover for

analyzer indication unit in the reverse

procedure. Turn ON the power supply SW.

The work will be completed if the analyzer

is normally worked.

(2) Display the moisture interference

compensation screen of the analyzer unit
(see “6.8 Maintenance mode”). Set the
dew point to 2

ЊC by using an electronic

cooler, and introduce bubbled N

2

or air

gas to the analyzer (shown on the figure).

(3) On the screen, select a desired Ch

(component) by pressing the

ENT

key,

adjust the value at right by pressing the

or the

key so that the value at left

falls within

±

10 (make it as close to 0 as

possible), and then press the

ENT

key to

memorize the value. (Exiting by “

ESC

cancels the adjustment.)

Or, selecting the “ALL” and pressing the

ENT

” key, zeroes all components

integrally.

(First, adjust all components by selecting
ALL and then perform fine adjustment
for components one by one using

and

keys.)

* If any components exceed the range of

0 to 10vol%, no adjustment can be
performed (No interference
compensation is required).