Yokogawa IR400 User Manual
Page 55

6 - 16
IM 11G02N01-01E
Remote start
Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the
remote start input closed for at least 1.5 seconds.
Remote start input
Open
Closed (keep at least 1.5 sec.)
Start Time
Cycle
Flow Time
ON/OFF
SUN
1
Zero
Ch1 Span
Ch2 Span
Ch3 Span
Ch4 Span
Ch5 Span
EX. time
ON
12:00
day
350 sec
350 sec
350 sec
350 sec
300 sec
300 sec
300 sec
(An example of “Ch1: through Ch5: enable”, as given in Item 6.2.4 “Auto Calibration
Components/range”)
Flow time
: Auto calibration
Sunday
12:00
Cycle
Zero
calibration
350sec
350sec
Ch1
Span
calibration
350sec
Ch2
Span
calibration
Ch3
Span
calibration
Ch4
Span
calibration
Monday
12:00
Tuesday
12:00
Example
In case where auto calibration is carried out at the above setting.
Ch5
Span
calibration
350sec
300sec
300sec
Replace-
ment time
300sec
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7days)
Flow time
: 60 to 900 sec
(initial value 300sec)
Setting range
When an auto calibration starts, the measurement screen automatically appears.
Any key operation other than key lock ON/OFF and “Stop Auto Calibration” (see
Item 6.4.2) is not permitted during auto calibration. “Stop Auto Calibration” cannot
be performed with the key lock to ON. To cancel auto calibration forcedly, set
the key lock to OFF and then execute “Stop Auto Calibration”.
Turn on the power again after it is turned off (including the case of power
failure) at the time set as the next start time in auto calibration, and then
repeat it in the set cycle.
Caution
•
•
•
Auto calibration status contact output is closed during auto calibration (including Ex. time), and is open in
other cases.