Yokogawa IR400 User Manual
Page 71

6 - 32
IM 11G02N01-01E
Reset input
So long as close, resetting lasts.
At the edge of changing from closing to opening, the average action restarts.
Close (hold at least 1.5 sec.)
Open
Average value reset
This mode is used to clear all average values O
2
correction average and O
2
average, and restarts
averaging. All average values are reset at a time. The indication value and output value is 0 ppm,
vol% or so at the time of the reset input (Refer to the average peripd).
Response time
The response time of the electrical system can
be changed.
Setting is available by components.
Note) It does not provide exact seconds for
the setting time, but it gives a guide
of the setting time.
The setting value can be modified as
requested by the customer.
Average period
It allows you to set an average period of the
average value of O
2
correction
and O
2
average.
It enables you to set an average time of 1 to 59
minutes (1-minute step) or 1 to 4 hours (1-hour
step).
Changing the setting resets the average value
of O
2
correction and O
2
average value.
(Pressing the
ENT
validates the resetting only
for components whose setting was changed.)