Factory page overview, Programming page overview – Watlow Series SD31 PID with Optional Countdown Timer User Manual
Page 20

Wa t l o w S e r i e s S D 3 1
■
1 8
■
C h a p t e r 3 K e y s a n d D i s p l a y s
Factory Page
[AMb] Ambient Temperature
[A;mn] Minimum Recorded Ambient Temperature
[A;ma] Maximum Recorded Ambient Temperature
[DSPL] Display Intensity
[A;Ot1] Output 1 Process Value
[rESt] Restore Factory Calibration
[Usr;r] Restore User Settings
[USr;S] Save User Settings
[dFLt] Default Parameters
[O;ty1] Output 1 Type
[O;ty2] Output 2 Type
[`S;Id] Software ID
[S;UEr] Software Version
[S;bld] Software Build Number
[PWr] Power Type
[`Sn-] Serial Number 1 (first four digits)
[`Sn_] Serial Number 2 (last four digits)
[tc;50] Thermocouple, 50mV
[tc;00] Thermocouple, 0mV
[tc;32] Thermocouple, 32°F
[`r;15] RTD, 15 ohm
[r;380] RTD, 380 ohm
[``U;1] Input Calibrate, 1.0 Volt
[``U;9] Input Calibrate, 9.0 Volt
[``A;4] Input Calibrate, 4.0 mA
[`A;16] Input Calibrate, 16.0 mA
[O1;1u] Output 1 Calibrate, 1.0 Volt
[O1;9u] Output 1 Calibrate, 9.0 Volt
[O1;4A] Output 1 Calibrate, 4.0 mA
[O1;16] Output 1 Calibrate, 16.0 mA
¯
Factory Page Overview
The Factory Page contains information on diagnostics,
calibration and restore-parameter functions.
To go to the Factory Page, press both the Up
¿ and
Down
¯ keys for about six seconds from the Home
Page.
•
Press the Down
¯ or Up ¿ keys to move through
the Factory Page parameters.
•
To view a parameter value, press and hold the
ß Key.
•
To change a parameter value, press and hold the
ß Key and use the Down ¯ or Up ¿ keys to
change the parameter value.
•
Press the Infinity Key
ˆ at any time to return to
the Home Page.
Note: Hardware configuration and programming selections deter-
mine what parameters appear on the Factory Page.
[FAct]
Programming Page Overview
The Programming Page determines what parameters
the user wants to appear on the Operations Page. Se-
lect a parameter for any of the 20 Programming Page
locations, P1 to P20. These now appear on the Opera-
tions Page. All 20 locations have parameters selected as
defaults.
To go to the Programming Page, hold down the Infinity
key
ˆ, then press the SET Key ß, and hold both
down for about six seconds.
•
Press the Down
¯ or Up ¿ keys to move through
the Programming Page parameters, P1-P20.
•
To view a parameter value, press and hold the
ß Key.
•
To change a parameter value, press and hold the
ß Key and use the Down ¯ or Up ¿ keys to
change the parameter value.
•
Press the Infinity Key
ˆ at any time to return to
the Home Page.
Note: The hardware configuration and programming selections will
also determine what parameters appear on the Operations Page. A
Programming Page selection will not appear on the Operations
Page if the parameter is not active.
[Prog]
Programming Page
[~~P1] Parameter Location 1 (48)*
[~~P2] Parameter Location 2 (49)
[~~P3] Parameter Location 3 (50)
[~~P4] Parameter Location 4 (51)
[~~P5] Parameter Location 5 (52)
[~~P6 Parameter Location 6 (53)
[~~P7] Parameter Location 7 (54)
[~~P8] Parameter Location 8 (55)
[~~P9] Parameter Location 9 (56)
[~P10] Parameter Location 10 (57)
[~P11] Parameter Location 11 (58)
[~P12] Parameter Location 12 (59)
[~P13] Parameter Location 13 (60)
[~P14] Parameter Location 14 (61)
[~P15] Parameter Location 15 (62)
[~P16] Parameter Location 16 (63)
[~P17] Parameter Location 17 (64)
[~P18] Parameter Location 18 (65)
[~P19] Parameter Location 19 (66)
[~P20] Parameter Location 20 (67)
¯
*Programming Page parameters P1 through P20 are