Section 3: Serial Communication Wiring

01/13

15

51297 Rev. 1

2.2.3.

CHECKING POWER SUPPLY / BATTERY VOLTAGE

1. Press the

TARE

key for three (3) seconds.

– The battery current or power supply voltage displays
8.3 VDC nominal.

2.2.4.

PROGRAMMING THE SLEEP FUNCTION

1. Press and hold the

BG/Net

key for three (3) seconds.

– The display will indicate

SLEEP

, and then display the current setting.

– This function serves to prolong battery life by turning off the Instrument.
– When there is no activity, the Sleep Mode activates according to the

programmed time frame.

– Activity is when weight is applied to the scale exceeding the balance setting,

pressing a key, and receiving a polling request on a communication port.

2. Press

MENU

.

3. Use the

Arrow Keys

to toggle between the available options.

– Selections in minutes include

OFF, 1, 2, 5, 10, 20, 30.

4. Press

ENTER

to save the setting.

2.3. Operating Procedures

1. Press and hold the

ON / OFF

Key for two (2) seconds.

– The Instrument will display “888888”, followed by “1234567890” character

display moving from right-to-left, then the revision of software.

– Once the warm-up is complete, the FB2255 displays the actual weight on the

scale.

2. To turn the FB2255 off, press the

ON / OFF

key for two (2) seconds.

The Zero. Tare and the AZT functions require the displayed weight to be stable

before these functions will operate.

The weight reading is stable if the variation in weight is less than the programmed

BAL RANGE

.