Section 5: Programming Configuration

04/13

36

51278 Rev. 2

8. rE.SAnP - Auto Re-sample Setting

.

Display

Descriptions

rE.SAnP no

Disable auto re-sample operation.

rE.SAnP YES

Enable auto re-sample operation. (Default)

Example:

1. Set rE.SAnP = YES.

2. Press

to into sample mode.

3. Press

to set the item number as 50. It is shown on the third line.

4. Place 5 lb item on scale weighing pan.

5. Wait for 5 seconds until the weight is stable. The piece weight shows 0.1 and

piece count shows 50.

6. Place 1 lb item on scale weighing pan.

7. At a stable weight, the scale will emit a short beep sound. Perform a re-

sample once. The Piece Weight will update.

8. Cycle power to the scale and enter the rE.SAnP setting again, change the

rE.SAnP = no.

9. Repeat steps 2 to 6.

10. At a stable weight, the scale will continue normal operation without a resample

process occurring.



10. tiNE - Time Setting

.

Display

Descriptions

tiNe no

Do not change time setting.

tiNe YES

Change time setting.

Note:

The time setting parameter always reverts back to “no”. When “YES” is selected, the
following parameters are available YEAr, NontH, dAY, hour, Nin, and SEC.