Section 5: Programming Configuration

5.1. INTRODUCTION

The Omega series scales are configured through internal programming parameters
and settings. Although accessible, the calibration parameters are restricted to
qualified service personnel. Changes to the calibration parameters will affect the
scales weighing and counting accuracy.

5.2. FACTORY DEFAULTS

The factory defaults are recommended to be loaded upon a first time installation or
instrument repair. To load the factory defaults, perform the following:

1. Press and hold any key (other than the On and Off keys) while powering on the

scale.

2.The display will indicate SEtUP

3.Press “1978” using the numeric keypad

4.The display will indicate SAvE, followed by a double beep, then the scale

will return to the weighing mode.

5.3. PROGRAMMING PARAMETERS

1. To access the programming parameters, press and hold any key while turning the

scale ON, SEtUP is shown on the display.

2. Press

key to toggle among the SEtUP, Prt, Con-F, rE.CAL, and CAL menus.

Use the

to enter selected menu.

Menu

Function

SEtUP

General Settings menu.

Prt

Printing Settings menu.

Con-F

Configuration Settings menu.

rE.CAL

External Scale Settings menu.

CAL

Calibration menu