Infrared Gas Analyzer

Maintenance

Teledyne

Analytical

Instruments

99

Figure 4-6: Cell Block Assembly

4.4 Optical Zero Adjustment

If the following operation is not performed correctly,
subsequent analysis may adversely be affected. This
operation should be performed by a trained technician or
the unit sent back to the factory for service.

The following adjustment is performed at reassembly after

removing and cleaning the sample cell.

1. Remove the top cover.
2. Allow dry N

2

or air to flow through the analyzer sample

gas inlet until the reading stabilizes. The sample gas is