Operation Model

7600

Teledyne Analytical Instruments

62

When a channel enabled for auto calibration is in the calibration

mode, a message indicating ZERO cal. or SPAN cal. will appear on the
measurement screen. For example, if channels 1 and 2 have the auto
calibration setting enabled, when the calibration event takes place, both
channels 1 and 2 will indicate ZERO cal. when the zero calibration is
being performed. Zero calibrations are performed first and all enabled
channels will zero simultaneously.

When the zero calibrations are finished, channel 1 will enter the

span calibration followed by channel 2.

3.5.4.4

R

EMOTE

S

TART

An auto calibration is always available by keeping the remote start

input closed for at least 1.5 seconds. The remote start can initiate an auto
calibration whether the parameter setting is ON or OFF.

3.5.4.5

F

ORCED

R

UN

/S

TOP OF

A

UTO

C

ALIBRATION

Auto calibration can be performed manually or forcibly stopped

while the calibration is performed.