Operation Model

7600

Teledyne Analytical Instruments

40

Where:

On: The value of the O

2

correction reference value (Value set by

application)

Os: Oxygen concentration (%)
Cs: Concentration of relevant measured component. Note that Os

does not exceed the O

2

limit value set in “Other Parameter”

See Section 3.6 Maintenance Mode.

The corrected sampling components are NO

X

, SO

2

and CO only.

O

2

correction concentration average value:

In the Ch (component) and O

2

average value where “CV/AV **” is

displayed as “ AV CO” in the component display, a value obtained by
averaging O

2

correction concentration value or O

2

average value in a

fixed time is output every 30 seconds.

Averaging time can be changed between 1 minute and 59 minutes

or 1 hour and 4 hours according to the average time settings (See
Section 3.5.7 Parameter Setting).

The set time is displayed as “1h” , for instance, in the range display.

* The measurement ranges of O

2

correction concentration value and O

2

correction concentration average value are the same as that of the
measuring components. Also, the measurement range of O

2

average

value is the same as that of O

2

.

3.3.2 Setting/Selection Screen

The setting/selection screen is configured as shown in Figure 3-5:

• In the status display area, the current status is displayed.
• In the message display area, messages associated with

operation are displayed.

• In the setting item and selection item display area, items or

values to be set are displayed, as required. To work in a
particular area, move the cursor to any item by using the UP,
DOWN and LEFT/RIGHT keys.