Operation Model

7600

Teledyne Analytical Instruments

54

• High Limit: sets the upper value (concentration) above which

the alarm will trigger. Has an on-screen display.

• Low Limit: sets a lower value (concentration) below which

the alarm will trigger. Has an on-screen display

• High or Low Limit: triggers an alarm when the concentration

is above the high limit or below the low limit.

• High-High: sets the upper value (concentration) above which

the alarm will trigger.

• Low-Low: sets a lower value (concentration) below which

the alarm will trigger.

• Power Failure: (Alarm 6 only) normally closed contacts open

when power is removed.

Configuring the alarms involves:

• Selecting the alarm type
• Assigning it to a channel
• Establishing setpoints (concentration alarms only)
• Enabling or disabling it

A channel can be configured for multiple alarms.
ON/OFF enables the alarm function if set at ON, or disables the

alarm function if set at OFF. The H-limit value cannot be set below the
L-limit value, and the L-limit value cannot be set above the H-limit
value.

If it is desired to set the H-limit value below the L-limit value

already stored in the memory, reduce the L-limit value beforehand, and
vice versa.

A typical on-screen display for a High alarm is shown below.
When the H-limit alarm occurs, the “H-alarm” message comes on

in the field of relevant channel. Similarly for L-alarm. There is no screen
display for the HH-alarm”, and “LL-alarm” but otherwise, they function
like the H and L limit alarms.