Infrared Gas Analyzer

Operation

Teledyne

Analytical

Instruments

55

CAUTION:

ALARMS ARE INACTIVE FOR 10 MINUTES AFTER
TURNING ON POWER.

3.5.3.1

C

ONFIGURING THE

A

LARMS

Note: To change an alarm setting, first set the alarm ON/OFF

setting to OFF, and then change the value.

1. From the measurement mode, press the MODE key to

display the user mode.

2. Point the cursor to “Alarm Setting” by pressing the UP or

DOWN key and press the ENT key.

3. In the “Alarm Setting” screen that appears, point the

cursor to the alarm number you want to set (1 through 6)
using the UP or DOWN key. Press the ENT key when the
cursor is on the alarm number you desire.