Infrared Gas Analyzer

Operation

Teledyne

Analytical

Instruments

73

interval (1) and (2)

,

the peak count alarm remains ON. At (2), the peak

count has fallen to 4 per hour so the peak alarm is turned OFF.

Like the hysteresis of the alarm setting , the hysteresis prevents

possible chattering when measured gas is fluctuated near the alarm
value.

CAUTION:

FOR 10 MINUTES AFTER THE POWER IS TURNED
ON, A PEAK ALARM COUNTING IS NOT
PERFORMED.

Releasing peak count alarm:

To release the peak count alarm, set the peak alarm to OFF.

Turning the peak alarm back ON again resets the count to 0.

3.5.7 Parameter Setting

This menu allows you to adjust parameter settings such as:

• Time
• Key lock
• Output hold
• Reset average output
• Response time
• Average period
• Backlight timer
• Enter maintenance mode

Description of setting items:

Current Time: Current year, month, date, day of the week, hour, and
minute setting in this order.

Note: The clock backup time is 2 days. If power is off for 2 days or

longer, the time setting will need to be set again.

Key Lock:

Sets with ON/OFF. When set to ON, all key
operations except key lock OFF are disabled.

Output Hold:

Sets whether calibration output is held or not, and
the holding value setting.