3.6.5 Moisture Interference Adjustment

This item is described in Section 4.5 Moisture Interference

Adjustment.

For the values listed in the middle column of the screen, the

moisture interference for each component is already offset. The values
in the right column are the interference compensation coefficients used.

To change the coefficient:

1. Move the

cursor to a desired channel by pressing the UP

or DOWN key, and then press the ENT key. The
selected interference compensation coefficient at the right
is highlighted.

2. Check that the gas for moisture interference

compensation is flowing. Change the moisture
interference compensation coefficient using the UP or
DOWN key. Adjust the value until the value in the left
column is near zero. Then press the ENT key to log
moisture interference compensation value.

Note: An interference compensation detector is not provided so if

the 1st range is beyond 0 to 10 vol %, no interference
adjustment can be performed (no need).

3.6.6 Output Adjustment Screen

In this menu you can adjust the analog output. First, connect a

digital multi meter to the output terminal corresponding to the output to
be adjusted. See the Appendix for terminal identification on the terminal
blocks. Adjust the value so that 4 mA or 0 V is output at zero and 20
mA or 1 V is output at span.