Maintenance Model

7600

Teledyne Analytical Instruments

102

Figure 4-9: Beam Adjustment Plate

8. Move the beam adjustment plate sidewise to change the

value displayed at (c) or (g).

9. Move the beam adjustment plate sidewise to change the

value displayed at (d) or (h).

10. Repeat the procedures in (6) to 9) to make all the

displayed values come close to 0 as possible within ±100
range.

Note: Adjust the beam adjustment plate which is the nearest to

the zero adjustment knob first, and sequentially.

11. After the optical balance adjustment, remount the top

cover of the analyzer unit, then carry out a moisture
interference compensation adjustment, and perform zero
and span calibrations.

Note: Before moving the beam adjustment plate, loosen the

detector set bolts (just enough to make the plate movable
for snug adjustment.

4.5 Moisture Interference Compensation Adjustment

If the following operation is not performed correctly,
subsequent analysis may adversely be affected. This