A.3 Optional Functions

O

2

correction: Correction of measured NO, SO

2

and CO

gas concentrations into values at reference
O

2

concentration

Correction

formula:

Cs

Os

-

21

On

-

21

C

×

=

C: Sample gas concentration after O

2

correction

Cs: Measured concentration of sample gas

Os: Measured O

2

concentration (Limit setting:

1 to 20% O

2

)

On: Reference O

2

concentration (value

changeable by setting. 0 to 19% O

2

)

Average value after O

2

correction and O

2

average value calculation:

The result of O

2

correction or instanta-

neous O

2

value can be output as an av-

erage value in the preset period of time.
The moving average method is used for
averaging in which sampling is carried out
at intervals of 30 seconds.

Output is updated every 30 seconds. It is

the average value in the determined period
of time just before the latest update.
Averaging time is settable within 1 to 59
minutes (in increments of 1 minute) or 1
to 4 hours (in increments of 1 hour).

Average value resetting: The above-mentioned output of average

value is started from the initial state by
opening the average value resetting input
terminals after short-circuiting for 1.5 sec-
onds or longer.

Output is reset by short-circuiting and re-

started by opening