Operation Model

7600

Teledyne Analytical Instruments

36

MODE key:

Used to switch the mode.

LEFT/RIGHT
Arrow key:

Used to change the selected item by moving the
cursor and numeral digit.

UP key:

Used to change the selected item by moving the
cursor and to increase a numeric value.

DOWN key:

Used to change the selected item by moving the
cursor and to decrease a numeric value.

ESC key:

Used to return to a previous screen or cancel the
setting midway.

ENT key:

Used for confirmation of selected item or value, and
for performing a calibration.

ZERO key:

Used for zero calibration.

SPAN key:

Used for spancalibration.

3.2 Display and Available Menus

The display on the Model 7600 has two modes: measurement mode

and user mode. To change between modes, press the MODE key.

In the measurement mode, the display indicates the composition of

the sample gas. Up to 12 channels of output are available although the
screen only shows 5 channels at a time. To see additional channels,
scroll down or up using the UP or DOWN keys.

Calibration is performed from the measurement mode by pressing

ZERO or SPAN. See Section 3.7 Calibration.