Appendix Model

7600

Teledyne Analytical Instruments

110

Analog Output: 12 points max.

4 to 20 mA DC or 0 to 1V DC, non-
isolated output

max.load

550Ω for 4 to 20 mA DC

min.load 100kΩ for 0 to 1VDC

Analog Input: 0 to 1V DC non-isolated input from

externally installed Zirconia type O

2

sensor.

Relay Contact Output: 1a contact (250V AC/2A, resistive load)

Instrument error, calibration error, range
identification, auto calibration status,
pump ON/OFF, peak alarm.

1c contact (250V AC/2A, resistive load

selectable 6 outputs). High/Low limit
alarm contact output. Power disconnection
alarm.

Contact

Input:

No-voltage contact (ON/0V, OFF/5V
DC, 5mA flowing at ON)
Remote range switch, auto calibration
remote start, remote holding, average
value resetting, pump ON/OFF

Isolated from the internal circuit with

photocoupler. Contact inputs are not iso-
lated from one another.

Other Output: Solenoid valve drive signal for automatic

calibration.

Transistor output (100mA or less)

Power Supply: Voltage rating: 100V to 240V AC

Allowable range: 85V to 264V AC
Frequency: 50Hz/60Hz

Power consumption: 250VA max.

Inlet:

Conforms

to

EN60320 Protection Class 1