Operation Model

7600

Teledyne Analytical Instruments

70

Note: During auto zero calibration, the keys are inoperable other

than key lock ON/OFF and “Stop Auto Zero Calibration.”

When the key lock is set to ON, even the “Auto Zero
Calibration Stop” is locked out. To stop “Auto Zero
Calibration” forcedly, set the key lock to OFF and then
execute “Auto Zero Calibration Stop” as detailed above.

3.5.6 Peak Alarm Setting

The optional peak alarm is triggered whenever the CO

concentration exceeds the upper limit value for a preset number of times.
The peak number is set from within the Peak Alarm Setting menu.

Note: Only instruments with the peak alarm option installed will

have this menu.

The following items are set within the peak alarm menu:

• OFF/ON status of peak alarm
• Alarm setpoint value
• Alarm count (number of peaks)
• Hysteresis

To adjust the peak alarm settings:

1. From the measurement mode, enter the user mode by

pressing the MODE key.

2. Point the cursor to “Setting of Peak Alarm” by pressing

the UP or DOWN key.