Operation Model

7600

Teledyne Analytical Instruments

68

3.5.5.2

F

ORCED RUN

/

STOP OF AUTO ZERO CALIBRATION

Like the auto calibration, the auto zero calibration can be

performed manually or forcibly stopped while the calibration is
performed.

3.5.5.2.1

M

ANUAL

E

XECUTION OF

A

UTO ZERO

C

ALIBRATION

To perform the auto calibration cycle manually:

1. From the measurement mode, press the MODE key to

display the user mode.

2. Point the cursor to “Setting of Auto Zero Calibration” by

pressing the UP or DOWN key and press the ENT key.

3. In the “Setting of Auto Zero Calibration” item selection

screen that appears, point the cursor to “Auto zero
Calibration Run” by pressing the

UP or DOWN

keys.

Press the ENT key.

4. “Run” is highlighted, displaying a message to confirm the

execution of auto calibration. Press the ENT key to
execute the auto calibration, or press the ESC key to
cancel.