Maintenance Model

7600

Teledyne Analytical Instruments

98

Figure 4-5: Detector Secure Nut

5. Using the furnished cell mounting tool, turn the window

fixture to the left and remove it from the cell. See Figure
4-6.

6. For cleaning the infrared transmission window and cell

inside surface, first eliminate coarse dust using a soft
brush and then wipe clean with a soft rag.

CAUTION:

THE WINDOW IS EASILY SCRATCHED. USE
EXTREME CARE WHEN PERFORMING THIS
OPERATION SO AS NOT TO DAMAGE IT.

7. After cleaning, reinstall the sample cell and perform an

optical zero adjustment (see Section 4.4) and moisture
interference compensation adjustment (see Section 4.5).