Infrared Gas Analyzer

Operation

Teledyne

Analytical

Instruments

57

Figure 3-6: Hysteresis Example for a High Limit Alarm


An alarm output is turned ON if measurement value exceeds the

upper limit value as shown in Figure 3-6. Once the alarm output has
been turned ON, it is not turned OFF until the process value falls below
the hysteresis width.

Note: The hysteresis setting is common to all alarms.

To set the hysteresis value:

1. From the “Alarm No. Selection” screen, point the cursor

to “Hysteresis” by pressing the UP or DOWN key then
press ENT.

2. In the “Hysteresis Value Setting” screen that appears,

enter hysteresis values using the UP or DOWN keys to
increment or decrement a digit. Use the SIDE key to
move to the next digit. When finished, press ENT.