Teledyne 7600 - Infrared gas analyzer - operating manual User Manual

Page 58

background image

Operation Model

7600

Teledyne Analytical Instruments

58

To close the “Hysteresis Setting” or cancel the mode midway, press

the ESC key. A previous screen will return.

3.5.4 Setting of Auto Calibration

3.5.4.1

A

UTO

C

ALIBRATION

Auto calibration is automatically carried out when zero and span

calibrations are set.

In this menu you can set when and how frequent the auto

calibration is performed. You also enable or disable the auto calibration
feature here.

Description of setting items:

Start Time: Setting at the first calibration (day of the week, hour,

minute)

Cycle:

A period between the start time of one calibration and

another (unit: hour/day)

Flow Time: The time required for replacement by calibration gas.

Time required for replacement of sample gas after the
calibration is completed (Set by calibration gas.)

ON/OFF: Enable/disable the auto calibration feature.

Note: Before changing the settings of auto calibration, set the

ON/OFF to OFF.

To change the settings:

3. From the measurement mode, press the MODE key to

display the user mode.

4. Point the cursor to “Setting of Auto Calibration” by

pressing the UP or DOWN key and press the ENT key.