beautypg.com

Maintenance – Teledyne 7600 - Infrared gas analyzer - operating manual User Manual

Page 93

background image

Infrared Gas Analyzer

Maintenance

Teledyne

Analytical

Instruments

93

Maintenance

4.1 Routine Maintenance

Aside from normal cleaning of the instrument and making sure

connections are tight and secure, the following maintenance items
should be carried out on a daily basis.

1. Zero and Span Calibration:

Perform a zero calibration. For the calibration
procedures, refer to Section 3.7.1 Zero Calibration.

Then, perform span calibration. For the calibration
procedures, refer to Section 3.7.2 Span Calibration.

2. Flowrate Check:

Sampling and purge gas flowrates are as follows:

Sampling gas flow: 0.5L/min 0.2 L/min
Purge gas flow:

About 1 L/min

Check and adjust if necessary every day.

4.2 Daily Check and Maintenance Procedures

Table 4.1 Troubleshooting and Maintenance

Item to be

checked

Phenomena Cause

Remedy

Display value

Display values are
low.

Display values are
high.

(1) Dust is
mixed in
sampling cell.

(1) Clean the
sampling cell.

Check
sampling
system
especially gas
filter.