3.5.4.5.1

M

ANUAL

E

XECUTION OF

A

UTO

C

ALIBRATION

To perform the auto calibration cycle manually:

1. From the measurement mode, press the MODE key to

display the user mode.

2. Point the cursor to “Setting of Auto Calibration” by

pressing the UP or DOWN key and press the ENT key.

3. In the “Setting of Auto Calibration” item selection screen

that appears, point the cursor to “Auto Calibration Run”
by pressing the

UP or DOWN

keys. Press the ENT

key.

4. “Run” is highlighted, displaying a message to confirm the

execution of auto calibration. Press the ENT key to
execute the auto calibration, or press the ESC key to
cancel.

3.5.4.5.2

F

ORCIBLY

E

ND AN

A

UTO

C

ALIBRATION

To abort an auto calibration event: