Infrared Gas Analyzer

Operation

Teledyne

Analytical

Instruments

89

3. Another screen appears which displays the current MIN

and MAX ranges established for that component. To
change the existing range move the cursor to the range
you want to modify using the UP or DOWN key, and
then press the ENT key.

Note: The value for range 1 and range 2 must fall within or be

equal to the MIN the MAX range and range 1 must be less
than range 2.

The number of ranges (Range num.) is 1 or 2.

4. Press the UP or DOWN key to change the value. Press

the SIDE key to select the next digit. When the decimal
point is highlighted, press the UP or DOWN key, and the
decimal point position can be changed.

5. When complete, press the ENT key to save the change.