3.

In the “Setting of Auto Zero Calibration” screen that
appears, point the cursor to any item you want to set by
pressing the

UP or DOWN

key. Press the ENT

key.

4. In the “Auto Zero Calibration Parameter Setting” screen

that appears, perform the value entry or the setting. For
the value entry or setting change, use the UP or DOWN
key. To change the setting, use the SIDE key to move the
cursor to the right.

5. After changing the setting, press the ENT

key to accept

the change. Auto zero calibration will now be carried out
using the entered setting value.

To close the “Setting of Auto Calibration” or cancel the mode

midway, press the ESC key. A previous screen will return.

The contacts for the auto zero calibration are closed during auto

zero calibration.

Auto Zero Calibration Example

Start Time

SUN

12:00

Cycle 12

hour

Flow Time

300

sec

ON/OFF ON


Figure 3-8 shows the auto zero calibration with the above settings.