Operation Model

7600

Teledyne Analytical Instruments

76

Auto calibration

Figure 3-11: Output Hold for Auto Calibration

External hold

Figure 3-12: Output Hold for External Hold

Screen display during hold

For Auto calibration and external holds, the “on Hold” message

will be shown blinking on the measurement screen.

In a manual calibration, the screen displays the process of

calibration so the “on Hold” message is not displayed even if the output
signal is held. However, “on Hold” message is displayed during the hold
extending time shown in Figure 3-10.

For both manual and auto calibration, if the calibration is cancelled

after the calibration gas has been directed to the analyzer, the hold
extending time will be performed.

There are two types of hold values that you can use: “current” or

“setting”. Current uses the existing value immediately before entering
output hold, while “setting” uses an arbitrary value that you set before
initiating the hold.

To change or set the value for a hold: