Response time (90%): 15 sec electrical response.

Within 60 seconds including replacement
time of sampling gas (with 0.5L/min gas
flow rate). Gas replacement time depends
on the number of components,and
measuring range.

A.5 Standard Requirements for Sample Gas

Flowrate: 0.5L / min ±0.2 L/ min

Temperature: 0 to 50°C

Pressure: 10 kPa (1.45 psi) or less (Gas outlet side

should be open to the atmospheric air.)

Dust:

100

g/Nm

3

in particle size of 1μm or less

Mist:

Unallowable

Moisture: Below saturation at 2°C (condensation un-

allowable).

Corrosive component: 1 ppm or less

Standard gas for calibration: Zero gas: Dry N

2

Span gas: Each sample gas having

concentration 90 to 100% of its measuring
range (recommended). Concentration
above 100% FS is unusable.

If an external zirconia O

2

sensor is

installed and calibration is carried out on
the same calibration gas line:

Zero gas: Dry air or atmospheric air

(provided without CO

2

sensor).

Span gas: For other than O

2

measurement,

each sample gas having concentration 90
to 100% of its measuring range.

For

O

2

measurement: O

2

gas of 1 to 2 vol

%.