Brooks, Mf series – Brooks Instrument Mfi Series User Manual
Page 41

4-3
Brooks
®
Mf Series
Section 4 Maintenance &
Troubleshooting
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
Table 4-1 Bench Troubleshooting
Trouble
Possible Cause
Check/Corrective Action
Actual flow overshoots controller setpoint
Anticipate potentiometer out of
Adjust anticipate potentiometer. Refer to Section 3-5.
by more than 5% full scale.*
adjustment
Output stays at zero and there is flow
Clogged sensor.
Clean sensor.Refer to cleaning, Section 4-2d.
through instrument
Output stays at zero regardless of setpoint/
Clogged control valve.*
Check TP3 with the setpoint at 100%. If the voltage is greater
flow and there is no flow through the
than 11V, disassemble and repair the control valve. Refer to
controller.
Section 4-4c.
Valve override input is grounded.*
Check valve override input (TB-2, Terminal 5).
Defective PC board.
Replace PC board. Refer to Section 4-4.
Output signal stays at +6.8 V or 26 mA
Valve stuck open or leaky.*
Clean and/or adjust control valve. Refer to cleaning procedure
regardless of setpoint and there is flow
and/or Section 4-4c.
through the instrument.
+15 V -28 Vdc applied to the valve
Check the valve override terminal (TB-2, Terminal 5).
override input.*
Defective PC board.
Replace PC board. Refer to Section 4-4.
Controller output signal follows setpoint at
Leaky control valve.
Disassemble and repair valve. Refer to Section 4-4c.
higher setpoints but will not go below 2% .*
Controller output signal follows setpoint at
Insufficient inlet pressure or pressure drop.
Adjust pressures, inspect in-line filters and clean/replace as
lower setpoints but does not reach
necessary.
full scale.*
Partially clogged sensor.
Check calibration. Refer to Section 3-4.
Partially clogged valve.
Disassemble and repair control valve. Refer to Section 4-4.
Valve out of adjustment.
Adjust valve. Refer to Section 4-4.
Valve guide spring failure.
Controller oscillates (see below).
Instrument grossly out of calibration. Flow is
Partially clogged sensor.
Clean sensor, refer to the cleaning procedure, Section 4-2d.
higher than desired.
Instrument grossly out of calibration. Flow is
Partially clogged restrictor.
Replace restrictor. Refer to Section 4-4.
lower than desired.
Instrument oscillates.
Pressure drop or inlet pressure excessive.*
Adjust pressures.
Oversized orifice.*
Check orifice size. Refer to Section 4-5.
Valve out of adjustment.*
Adjust valve. Refer to Section 4-4c.
Anticipate potentiometer out of adjustment.
Adjust anticipate potentiometer. Refer to Section 3-4.
Faulty pressure regulator.
Check regular capacity.
Defective PC board.
Replace PC board. Refer to Section 4-4.
*For controller only (Models Mf50/51/53).