TFX Error Codes

Revised 2-22-2002

Code Number

Description

Correction

Warnings

0001

Serial number not present

Hardware serial number has become inoperative – system
performance will not be influenced.

0010

Signal Strength is below Signal
Strength Cutoff entry

Low signal strength is typically caused by one of the following:

Empty

pipe

Improper programming/incorrect values

Improper transducer spacing

Non-homogeneous pipe wall

0011

Measured Speed of Sound the in the
liquid is greater than 10% different than
the value entered during meter setup

Verify that the correct liquid was selected in the BASIC menu.
Verify that pipe size parameters are correct.

0020

Heat Flow Units of measure have been
selected and an RTD module has not
been installed

Verify that RTD Module PN D020-1045-106 has been installed in
one of the I/O meter slots. Verify that OUTPUT1 or OUTPUT 2 has
been configured for RTD measurements.

Class C Errors

1001

System tables have changed

Initiate a meter RESET by cycling power or by selecting SYSTEM
RESET in the SEC MENU.

1002

System configuration has changed

Initiate a meter RESET by cycling power or by selecting SYSTEM
RESET in the SEC MENU.

Class B Errors

3001

Invalid hardware configuration

Upload corrected file

3002

Invalid system configuration

Upload corrected file

3003

Invalid strategy file

Upload corrected file

3004

Invalid calibration data

Recalibrate the system

3005

Invalid speed of sound calibration data

Upload new data

3006

Bad system tables

Upload new table data

3007

Data Logger is off or not present

If desired, insert data logger and configure within the Datalog
Operations Menu. If logger is not present, configure I/O port for no
logger.

3010

One or more channels are not
responding (Multi-channel meters only)

Display indicates which secondary units are not communicating
with Master meter. Verify wiring, configuration and address of
secondary instrument.

3011

All channels are not responding (Multi-
channel meters only)

Verify wiring, configuration and address of secondary instruments.

Class A Errors

4001

Flash memory full

Return unit to factory for evaluation