Part 2 - transducer positioning – Dynasonics TFXM Series Ultrasonic Multi-Channel Flow Meter User Manual
Page 25

Rev. 8/02
- 2.10 -
TFXM
PIPE OD -- Pipe Outside Diameter Entry
ENGLSH (Inches)
METRIC (Millimeters)
Enter the pipe outside diameter in inches if ENGLSH was se-
lected as UNITS; in millimeters if METRIC was selected.
PIPE WT -- Pipe Wall Thickness Entry
ENGLSH (Inches)
METRIC (Millimeters)
Enter the pipe wall thickness in inches if ENGLSH was selected
as UNITS; in millimeters if METRIC was selected.
PIPE MAT -- Pipe Material Selection
CARBON S - Carbon Steel
STAINLES - Stainless Steel
CAST IRO - Cast Iron
DUCTILE - Ductile Iron
COPPER - Copper
PVC - Polyvinylchloride
PVDF LOW -
Low Density Polyvinylidene Flouride
PVDF HI -
High Density Polyvinylidene Flouride
ALUMINUM - Aluminum
ASBESTOS - Asbestos Cement
FIBERGLA - Fiberglass
OTHER
This list is provided as an example. Additional materials are be-
ing added continuously. Select the appropriate pipe material
from the list or select OTHER if the material is not listed.
PIPE SS -- Speed of Sound in the Pipe Material
ENGLSH (Feet per Second)
METRIC (Meters per Second)
Allows adjustments to be made to the speed of sound in the
pipe wall. If the UNITS value was set to ENGLSH, the entry is
in FPS (feet per second). METRIC entries are made in MPS
PART 2 - TRANSDUCER POSITIONING
Pipe O.D. Entry
Pipe Wall Entry
Pipe Material
Entry
Pipe Sound
Speed Entry